Please use this identifier to cite or link to this item: https://hdl.handle.net/20.500.11851/2787
Title: Integrated Hifu Drive System on a Chip for Cmut-Based Catheter Ablation System
Authors: Farhanieh, Omid
Sahafi, Ali
Roy, Rupak Bardhan
Ergün, Arif Şanlı
Bozkurt, Ayhan
Keywords: Capacitive micromachined ultrasound transducer (CMUT)
catheter ablation system
continuous wave high-voltage driver
high Intensity Focused Ultrasonic (HIFU)
phase locked loop
real-time HIFU thermal image
Publisher: Institute of Electrical and Electronics Engineers Inc.
Source: Farhanieh, O., Sahafi, A., Roy, R. B., Ergun, A. S., and Bozkurt, A. (2017). Integrated HIFU drive system on a chip for CMUT-based catheter ablation system. IEEE transactions on biomedical circuits and systems, 11(3), 534-546.
Abstract: Conventional High Intensity Focused Ultrasound (HIFU) is a therapeutic modality which is extracorporeally administered. In applications where a relatively small HIFU lesion is required, an intravascular HIFU probe can be deployed to the ablation site. In this paper, we demonstrate the design and implementation a fully integrated HIFU drive system on a chip to be placed on a 6 Fr catheter probe. An 8-element capacitive micromachined ultrasound transducer (CMUT) ring array of 2 mm diameter has been used as the ultrasound source. The driver chip is fabricated in 0.35 mu m AMS high-voltage CMOS technology and comprises eight continuous-wave (CW) high-voltage CMUT drivers (10.9 ns and 9.4 ns rise and fall times at 20 V-pp output into a 15 pF), an eight-channel digital beamformer (8-12 MHz output frequency with 11.25 degrees phase accuracy) and a phase locked loop with an integrated VCO as a tunable clock source (128-192 MHz). The chip occupies 1.85 x 1.8 mm(2) area including input and output (I/O) pads. When the transducer array is immersed in sunflower oil and driven by the IC with eight 20 Vpp CW pulses at 10 MHz, real-time thermal images of the HIFU beam indicate that the focal temperature rises by 16.8 degrees C in 11 seconds. Each HV driver consumes around 67 mW of power when driving the CMUT array at 10 MHz, which adds up to 560 m W for the whole chip. FEM based analysis reveals that the outer surface temperature of the catheter is expected to remain below the 42 degrees C tissue damage limit during therapy.
URI: https://hdl.handle.net/20.500.11851/2787
https://ieeexplore.ieee.org/document/7882666/
ISSN: 1932-4545
Appears in Collections:Elektrik ve Elektronik Mühendisliği Bölümü / Department of Electrical & Electronics Engineering
PubMed İndeksli Yayınlar Koleksiyonu / PubMed Indexed Publications Collection
Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection

Show full item record



CORE Recommender

SCOPUSTM   
Citations

12
checked on Dec 21, 2024

WEB OF SCIENCETM
Citations

23
checked on Dec 21, 2024

Page view(s)

72
checked on Dec 23, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.