Please use this identifier to cite or link to this item: https://hdl.handle.net/20.500.11851/2796
Title: An Integrated Beamforming Driver for CMUT based Ultrasound Catheter Ablation System
Authors: Yetik, Hasan
Ergün, Arif Şanlı
Bozkurt, Ayhan
Keywords: Component
beamforming
beamforming driver
HIFU
VLSI
CMUT
Publisher: IEEE
Source: Yetik, H., Ergun, A. S., and Bozkurt, A. (2014, September). An integrated beamforming driver for CMUT based ultrasound catheter ablation system. In 2014 IEEE International Ultrasonics Symposium (pp. 1532-1535). IEEE.
Abstract: In CMUT based ultrasound catheter ablation systems, transducer arrays and their drivers are to be kept as small as possible due to size restrictions, while maintaining a reasonable number of channels for proper beamforming. In addition, these systems should be highly integrated with minimal power consumption for flexibility and functionality. Despite the size restrictions on these systems, such problems can be overcome by ASICs. In this work, we present a simple 8 channel high-frequency, high-voltage driver, together with its beamforming and control circuitry designed with AMS's H35 high voltage process.
URI: https://hdl.handle.net/20.500.11851/2796
https://ieeexplore.ieee.org/document/6932306
ISBN: 978-1-4799-7049-0
ISSN: 1948-5719
Appears in Collections:Elektrik ve Elektronik Mühendisliği Bölümü / Department of Electrical & Electronics Engineering
Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection

Show full item record



CORE Recommender

Page view(s)

52
checked on Nov 18, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.