Please use this identifier to cite or link to this item: https://hdl.handle.net/20.500.11851/5904
Full metadata record
DC FieldValueLanguage
dc.contributor.authorJafarov, A. A.-
dc.contributor.authorErbay, F. K.-
dc.contributor.authorAlizadeh, C. A.-
dc.contributor.authorDemir, Teyfik-
dc.date.accessioned2021-09-11T15:20:41Z-
dc.date.available2021-09-11T15:20:41Z-
dc.date.issued2019en_US
dc.identifier.issn0005-2523-
dc.identifier.urihttps://hdl.handle.net/20.500.11851/5904-
dc.description.abstractIn order to determine the structural advantage of new H beam implant, pull out tests were conducted for perforated and non-perforated models of H beam implant until the" H" beam implants pulled from the polyurethane blocks which were used as a bone model. Tests were performed with a total of 12 H-beam implants with 6 perforated and 6 non perforated implant. Polyurethane (PU) blocks were produced containing polyol and polyisocyanate. The polymerization process act as fusion of soft tissue with H-beam implants. The mean maximum pull-out force for the perforated H-beam implant was 3215.72 ± 559.06 N and 2643.68 ± 718.94 N for non-perforated implants. Although the pull-out strength of perforated implants was 20.2% higher than H-beam implants, no statistically significant difference was found between them. The study showed that a perforated implant can accelerate the healing of femoral neck fracture and the stability of this implant is higher than the stability of conventional implants. © 2019 Ministry of Health. All rights reserved.en_US
dc.language.isoazen_US
dc.publisherMinistry of Healthen_US
dc.relation.ispartofAzerbaijan Medical Journalen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectFemoral neck fracturesen_US
dc.subjectPerforated implanten_US
dc.subjectPull-out testen_US
dc.titlePull Out Test Evaluation of the New Perforated Implant Designed for Osteosynthesis of Femoral Neck Fractureen_US
dc.title.alternativeBud sümüyü boynu siniqlarinin osteosintezi üçün tövsiy? edil?n yeni d?likli implantin ç?kib-çixarma testi il? qiym?tl?ndirilm?sien_US
dc.typeArticleen_US
dc.departmentFaculties, Faculty of Engineering, Department of Mechanical Engineeringen_US
dc.departmentFakülteler, Mühendislik Fakültesi, Makine Mühendisliği Bölümütr_TR
dc.identifier.volume2019en_US
dc.identifier.issue4en_US
dc.identifier.startpage111en_US
dc.identifier.endpage115en_US
dc.identifier.scopus2-s2.0-85078350497en_US
dc.institutionauthorDemir, Teyfik-
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.identifier.scopusqualityQ4-
item.openairetypeArticle-
item.languageiso639-1az-
item.grantfulltextnone-
item.fulltextNo Fulltext-
item.openairecristypehttp://purl.org/coar/resource_type/c_18cf-
item.cerifentitytypePublications-
Appears in Collections:Makine Mühendisliği Bölümü / Department of Mechanical Engineering
Scopus İndeksli Yayınlar Koleksiyonu / Scopus Indexed Publications Collection
Show simple item record



CORE Recommender

Page view(s)

116
checked on Dec 23, 2024

Google ScholarTM

Check





Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.